Free-air ionization chamber, FAC-IR-300, designed for medium energy X-ray dosimetry

The primary standard for X-ray photons is based on parallel-plate free-air ionization chamber (FAC). Therefore, the Atomic Energy Organization of Iran (AEOI) is tried to design and build the free-air ionization chamber, FAC-IR-۳۰۰, for low and medium energy X-ray dosimetry. The main aim of the present work is to investigate specification of the FAC-IR-۳۰۰ ionization chamber and design it. FAC-IR-۳۰۰ dosimeter is composed of two parallel plates, a high voltage (HV) plate and a collector plate, along with a guard electrode that surrounds the collector plate. The guard plate and the collector were separated by an air gap. For obtaining uniformity in the electric field distribution, a group of guard strips was used around the ionization chamber. These characterizations involve determining the exact dimensions of the ionization chamber by using Monte Carlo simulation and introducing correction factors.

مطالب مرتبط

نظرات یک دیدگاه جدید ارسال کنید.

برای ثبت نظر ابتدا باید وارد سایت شوید !