Free-air ionization chamber, FAC-IR-300, designed for medium energy X-ray dosimetry
- 12/آبان/1402
- مقالات ژورنالی انگلیسی
- نویسنده : آزمایشگاه میکرودزیمتری نوترون
- 11 بازدید
- 0 نظر
The primary standard for X-ray photons is based on parallel-plate free-air ionization chamber (FAC). Therefore, the Atomic Energy Organization of Iran (AEOI) is tried to design and build the free-air ionization chamber, FAC-IR-۳۰۰, for low and medium energy X-ray dosimetry. The main aim of the present work is to investigate specification of the FAC-IR-۳۰۰ ionization chamber and design it. FAC-IR-۳۰۰ dosimeter is composed of two parallel plates, a high voltage (HV) plate and a collector plate, along with a guard electrode that surrounds the collector plate. The guard plate and the collector were separated by an air gap. For obtaining uniformity in the electric field distribution, a group of guard strips was used around the ionization chamber. These characterizations involve determining the exact dimensions of the ionization chamber by using Monte Carlo simulation and introducing correction factors.